MF06 – 3.32 Installation, Operation & Maintenance Instructions 1000°C - 1200°C Furnaces types AAF, GSM, VMF This manual is for th
AAF, GSM, VMF 10 MF06 – 3.32 Switch the furnace ON and heat to 900°C without interruption, and then soak for 1 hour. Some smoke may be observed
AAF, GSM, VMF MF06 3.32 11 5.8 VMF 10/6 Element Replacement See section 5.2 - wearing a face mask is recommended. The heating elements are va
AAF, GSM, VMF 12 MF06 – 3.32 6.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON The heating element has failed Che
AAF, GSM, VMF MF06 3.32 13 F1 N F2 L1 PE transformer Filter (if fitted) F2 Instrument Switch 7.0 CIRCUIT DIAGRAMS EMC Filters (if
AAF, GSM, VMF 14 MF06 – 3.32 7.4 3-phase “universal” wiring Fuses F1 are always present in this wiring design. Fuses F2 could be
AAF, GSM, VMF MF06 3.32 15 8.0 FUSES & POWER SETTINGS 8.1 Fuses F1-F3: Refer to the circuit diagrams. F1 Internal supply fuses Fitted if
Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: [email protected] The pr
AAF, GSM, VMF 2 MF06 – 3.32 1.0 DESCRIPTIONS The chamber furnaces covered by this manual all have some specialist feature as follows. 1.1 AAF mo
AAF, GSM, VMF MF06 3.32 3 2.0 INSTALLATION 2.1 Unpacking & Handling When unpacking or moving the furnace always lift it by its base. Never
AAF, GSM, VMF 4 MF06 – 3.32 L3 Black to phase 3 except 2-phase N Light Blue to neutral except delta PE Green/Yellow to earth (grou
AAF, GSM, VMF MF06 3.32 5 3.0 OPERATION The instructions for operating the temperature controller are given in a separate manual. If the fu
AAF, GSM, VMF 6 MF06 – 3.32 Any metal object used to probe into the furnace chamber while the furnace is connected to the supply must be earthed
AAF, GSM, VMF MF06 3.32 7 4.0 MAINTENANCE 4.1 General Maintenance Preventive rather than reactive maintenance is to be preferred. The type and f
AAF, GSM, VMF 8 MF06 – 3.32 5.0 REPAIRS & REPLACEMENTS 5.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is di
AAF, GSM, VMF MF06 3.32 9 5.5 Thermocouple Replacement Disconnect the furnace from the supply, and remove the furnace back panel. Make a note