MF48 - 1.18
Installation, Operation &
Maintenance Instructions
1200°C G-range Tube furnaces
Models GHA, GVA, GHC, GVC
This manual is for the guidance of operators of the above Carbolite products
and should be read before the furnace is connected to the electricity supply.
Manuals are supplied separately for the furnace controller
(and overtemperature controller when fitted).
Please read the controller manuals before operating the furnace.
CONTENTS
Section page
1.0 Introduction 2
2.0 Installation 3
3.0 Operation 7
4.0 Maintenance 10
5.0 Repairs & Replacements 11
6.0 Fault Analysis 14
7.0 Circuit Diagrams 15
8.0 Fuses & Power Settings 19
9.0 Specifications 21
Vue de la page 0
1 2 3 4 5 6 ... 23 24

Résumé du contenu

Page 1 - Maintenance Instructions

MF48 - 1.18 Installation, Operation & Maintenance Instructions 1200°C G-range Tube furnaces Models GHA, GVA, GHC, GVC This manual is f

Page 2 - 2 MF48 -1.18

G range 10 MF48 -1.18 4.0 MAINTENANCE 4.1 General Maintenance No routine maintenance is required. The outer surfaces may be cleaned with a damp c

Page 3 - 2.0 INSTALLATION

G range MF48 -1.18 11 5.0 REPAIRS & REPLACEMENTS 5.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is disconnect

Page 4 - 4 MF48 -1.18

G range 12 MF48 -1.18 separate the cylindrical body from the base, or remove some guards under the protruding ends of the body. For the versatile m

Page 5

G range MF48 -1.18 13 Check that the furnace is controlling properly to rule out the possibility that the element failed because of a fault in the

Page 6 - 6 MF48 -1.18

G range 14 MF48 -1.18 6.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON  The heating element has failed  Check also

Page 7 - 3.0 OPERATION

G range MF48 -1.18 15 7.0 CIRCUIT DIAGRAMS For standard furnaces -/600 upwards, and -/450 if ordered 3-phase, see 7.7; for 3-zone see also 7.6. 7

Page 8 - 8 MF48 -1.18

G range 16 MF48 -1.18 3-phase without neutral (delta) In most instances this diagram does not apply. See 7.7. It may apply if non-standard models a

Page 9 - MF48 -1.18 9

G range MF48 -1.18 17 7.7 3-phase “universal” wiring (models -/600 and over, & /-450 3-phase) Fuses F1 are always present in t

Page 10 - 4.0 MAINTENANCE

G range 18 MF48 -1.18 7.8 3-Zone Control – Broadcast Comms HD HE HF centre (master) HD HE HF end (slave) HD HE HF end (slave)

Page 11 - MF48 -1.18 11

G range MF48 -1.18 19 8.0 FUSES & POWER SETTINGS 8.1 Fuses F1-F3: Refer to the circuit diagrams. F1 Internal supply fuses Fitted to al

Page 12 - 12 MF48 -1.18

G range 2 MF48 -1.18 1.0 INTRODUCTION 1.1 Products Covered This manual contains installation and operating instructions for the single-zone tube

Page 13 - MF48 -1.18 13

G range 20 MF48 -1.18 8.2 Power Settings & Adjustment The furnace control system incorporates electronic power limiting. The power limit param

Page 14 - 6.0 FAULT ANALYSIS

MF48 -1.18 21 9.0 SPECIFICATIONS 9.1 Models Covered by this Manual GHA models – horizontal single-zone tube furnaces with controls in integral ba

Page 16

MF48 -1.18 23

Page 17

Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: [email protected] The

Page 18 - 18 MF48 -1.18

G range MF48 -1.18 3 2.0 INSTALLATION 2.1 Unpacking & Handling When unpacking or moving the furnace always lift it by its base or by both en

Page 19 - USES & POWER SETTINGS

G range 4 MF48 -1.18 Work Tubes and Fittings Tubes of various diameters and lengths may be fitted. Type C: minimum length for the furnace Type D:

Page 20 - 20 MF48 -1.18

G range MF48 -1.18 5 2.5 Electrical Connections Connection by a qualified electrician is recommended. All furnaces covered by this manual may be

Page 21 - 9.0 SPECIFICATIONS

G range 6 MF48 -1.18 Models -/600 to -/1200 (and -/450 3-phase) These models are all made to a “universal” wiring scheme, whereby the customer may,

Page 22 - MF48 -1.18

G range MF48 -1.18 7 3.0 OPERATION The instructions for operating the temperature controller are given in a separate manual. If the furnace is f

Page 23 - MF48 -1.18 23

G range 8 MF48 -1.18 Set the “setpoint” of the end zone controllers to zero. Alternatively, a small temperature difference (offset) can be created

Page 24

G range MF48 -1.18 9 to the progressive shrinkage of the insulation materials. Cracks are not usually detrimental to the functioning or the safety