MF04 – 3.29 Installation, Operation & Maintenance Instructions 1200°C - 1300°C Chamber Furnaces GPC models This manual is for the gui
GPC 10 MF04 3.29 Disconnect the furnace from the electrical supply and remove the back panel. It is advisable to make a written plan showing a
GPC MF04 3.29 11 6.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON The heating element has failed Check also that
GPC 12 MF04 3.29 7.0 CIRCUIT DIAGRAMS In all GPC models the safety switch in the diagram is a positive-break door switch. EMC Filters (if fi
GPC MF04 3.29 13 7.3 3-phase without neutral (delta) 7.4 3-phase without neutral (star – e.g. 380 to 415V)
GPC 14 MF04 3.29 7.6 3-phase “universal” wiring Fuses F1 are always present in this wiring design. Fuses F2 could be absent in s
GPC MF04 3.29 15 8.0 FUSES & POWER SETTINGS 8.1 Fuses F1-F3: Refer to the circuit diagrams. F1 Internal supply fuses Fitted if supply c
Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: [email protected] The p
GPC 2 MF04 3.29 1.0 INTRODUCTION 1.1 GPC models This manual covers both the “laboratory” and “industrial” GPC furnaces. The laboratory model
GPC MF04 3.29 3 2.0 INSTALLATION 2.1 Unpacking & Handling Laboratory Models. When unpacking or moving the furnace always lift it by its ba
GPC 4 MF04 3.29 2.5 Electrical Connections Connection by a qualified electrician is recommended. All GPC models have three internal circuits
GPC MF04 3.29 5 3.0 OPERATION The instructions for operating the temperature controller are given in a separate manual. If the furnace is f
GPC 6 MF04 3.29 3.4 Atmospheres When an optional gas inlet is fitted there is a label near the inlet saying "INERT GAS ONLY". In prac
GPC MF04 3.29 7 4.0 MAINTENANCE 4.1 General Maintenance Preventive rather than reactive maintenance is to be preferred. The type and frequency d
GPC 8 MF04 3.29 5.0 REPAIRS & REPLACEMENTS 5.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is disconnected f
GPC MF04 3.29 9 5.5 Thermocouple Replacement Disconnect the furnace from the supply, and remove the furnace back panel. Make a note of the th